TEM Online Total Nitrogen Analyzer
In the potassium persulfate digestion UV spectrophotometric method, nitrogen compounds are oxidized to nitrate ions through high-temperature digestion using persulfate. Under acidic conditions, the resulting change is measured by UV absorbance. Turbidity effects are corrected during analysis, and the total nitrogen content is calculated based on the measured absorbance.
Product Application
● Monitoring at the inlet and outlet of wastewater treatment plants
● Effluent discharge points of industrial facilities and other pollution sources
Key words:
Classification:
- 产品描述
- Product Features
- Performance Index
- Field Application
-
TEM在线总氮分析仪采用碱性过硫酸钾消解紫外光分光光度法,与实验室的分析有很好的相关性,整体运行成本低,测量精准,无二次污染。
应用领域:
• 水厂:污水厂及自来水厂进出水口
• 各企业排污口
• 湖泊、河流水质监测站
-
- Measurement method complies with the national standard GB 11894-89, showing strong correlation with laboratory analysis.
- Innovative structural design provides lower failure rate, reduced maintenance, lower reagent consumption, and enhanced cost performance.
- Unique valve assembly allows flexible switching among multiple channels, delivering easy maintenance and extended service life.
- High-precision optical system ensures accurate reagent dosing, eliminating volumetric errors caused by wear of peristaltic pumps and tubing, while enabling precise micro-reagent metering.
- Multiple signal output interfaces to accommodate diverse system integration requirements.
- Equipped with robust quality control functions, including automatic calibration and standard solution verification, fully compliant with the latest industry standards.
-
Parameters
Details
Measuring Range
0–10 mg/L; 0–30 mg/L; 0–50 mg/L; 0–150 mg/L, extendable
Repeatability
≤ 2%
Indication Error
±10% (TN < 4 mg/L)
±8% (4 mg/L ≤ TN < 10 mg/L)
±5% (10 mg/L ≤ TN < 16 mg/L)
±5% (TN ≥ 16 mg/L)Actual Sample Accuracy
≤ 0.2 mg/L (TN < 2 mg/L)
≤ 10% (TN ≥ 2 mg/L)Quantification Limit
≤ 0.3 mg/L
24h Low Concentration Drift
≤ 0.05 mg/L
24h High Concentration Drift
≤ 1%
Measurement Cycle
Minimum cycle ≤ 35 min
Calibration
Supports online automatic calibration
Standard Solution Verification
Supports automatic verification with standard solutions
Automatic Cleaning
Automatic pipeline cleaning after each measurement
Reagent Consumption
0.7 mL/time/reagent
Output Interfaces
4–20 mA, RS485 with Modbus protocol
Power Supply
220 ± 22 VAC, 50 ± 2.5 Hz
Power Consumption
Average 50 W; peak < 300 W
Dimensions
420 × 600 × 250 mm
Ambient Temperature
5–40°C
Related Information
Contact us
Tengine is ready to help with quotes, technical issues, and all your questions. Simply fill out our form for a prompt response.
*Please ensure all information is accurately provided to enable us to respond to you promptly.
Tengine Innovation (Beijing) Instruments Inc.
Email: contact@tengine.com.cn
WhatsApp: +86 133 3109 8301
Address: Floor 6, Building 8, Yard 1,
Nongda South Road, Haidian District, Beijing
Copyright©2022 Tengine Innovation (Beijing) Instruments Inc.
Copyright©2022 Tengine Innovation
(Beijing) Instruments Inc.